One-year analysis of Elekta CBCT image quality using NPS and MTF.
Ontology highlight
ABSTRACT: The image quality (IQ) of imaging systems must be sufficiently high for image-guided radiation therapy (IGRT). Hence, users should implement a quality assurance program to maintain IQ. In our routine IQ tests of the kV cone-beam CT system (Elekta XVI), image noise was quantified by noise standard deviation (NSD), which was the standard deviation of CT numbers measured in a small area in an image of an IQ test phantom (Catphan), and the high spatial resolution (HSR) was evaluated by the number of line-pairs (LPN) visually recognizable in the image. We also measured the image uniformity, the low contrast resolution, and the distances of two points for geometrical accuracy. For this study, we did an additional evaluation of the XVI data for 12 monthly IQ tests by using noise power spectrum (N
SUBMITTER: Nakahara S
PROVIDER: S-EPMC5690923 | biostudies-literature | 2016 May