Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system.
Ontology highlight
ABSTRACT: The dose measurements of the small field sizes, such as conical collimators used in stereotactic radiosurgery (SRS), are a significant challenge due to many factors including source occlusion, detector size limitation, and lack of lateral electronic equilibrium. One useful tool in dealing with the small field effect is Monte Carlo (MC) simulation. In this study, we report a comparison of Monte Carlo simulations and measurements of output factors for the Varian SRS system with conical collimators for energies of 6 MV flattening filter-free (6 MV) and 10 MV flattening filter-free (10 MV) on the TrueBeam accelerator. Monte Carlo simulations of Varian's SRS system for 6 MV and 10 MV photon energies with cones sizes of 17.5 mm, 15.0 mm, 12.5 mm, 10.0 mm, 7.5 mm, 5.0 mm, and 4.0 mm were performe