Dosimetric verification and quality assurance of running-start-stop (RSS) delivery in tomotherapy.
Ontology highlight
ABSTRACT: The purpose of this study was to evaluate the dosimetric profiles and delivery accuracy of running-start-stop (RSS) delivery in tomotherapy and to present initial quality assurance (QA) results on the accuracy of the dynamic jaw motion, dosimetric penumbrae of the RSS dynamic jaw and the static jaw were measured by radiographic films. Delivery accuracy of the RSS was evaluated by gamma analysis on film measurements of 12 phantom plans. Consistency in the performance of RSS was evaluated by QA procedures over the first nine months after the installation of the feature. These QA were devised to check: 1) positional accuracy of moving jaws; 2) consistency of relative radiation output collimated by discrete and continuously sweeping jaws; 3) consistency of field widths and profiles. In the lon