ABSTRACT: Purpose
To compare the accuracy and speed of using the computerized MatchedFlicker software program (EyeIC Inc, Narberth, Pennsylvania, USA) to evaluate glaucomatous optic disc change against the traditional gold standard of manually examining stereoscopic disc photographs.Design
A prospective evaluation of diagnostic technology.Methods
Two resident ophthalmologists and 1 glaucoma fellow at the University of Florida independently evaluated 140 image pairs from 100 glaucomatous/ocular hypertensive patient eyes using a handheld stereo viewer and the MatchedFlicker program. Fifty had progression to glaucoma as determined by the Ocular Hypertension Treatment Study (OHTS) Optic Disc Reading Group and the OHTS Endpoint Committee in the OHTS, and 50 more had photographs ta