High variability in serum estradiol measurements in men and women.
Ontology highlight
ABSTRACT: To reduce the variability in estradiol (E2) testing and to assure better patient care, standardization of E2 measurements has been recommended. This study aims to assess the accuracy and variability of E2 measurements performed by 11 routine immunological methods and 6 mass spectrometry methods using single donor serum materials and to compare the results to a reference method. The contribution of calibration bias, specificity or matrix effects, and imprecision on the overall variability of individual assays was evaluated. This study showed substantial variability in serum E2 measurements in samples from men and pre- and post-menopausal women. The mean bias across all samples, for each participant, ranged between -2.4% and 235%, with 3 participants having a mean bias of over 100%. The data