Evaluation of gantry rotation overrun in axial CT scanning.
Ontology highlight
ABSTRACT: The purpose of this study was to develop and evaluate a simple method to assess gantry rotation overrun in a single axial CT scanning. The exposure time in the axial scanning was measured at selected nominal rotation times (400, 700, and 1000 ms) using a solid-state detector, the RTI's CT dose profiler (CTDP). CTDP was placed at the isocenter and the radiation dose rate signal (profile) was recorded. Subsequently, the full width of this profile was determined as the exposure time (Taxial). Next, CTDP was positioned on the inner cover of the gantry with a sheet of lead (1 mm thick) placed on top of the detector. Gantry rotation time (Thelical) was determined by the time between two successive radiation peaks during continuous helical scanning. The gantry overrun time (Toverrun) is, thus, de