Evaluation of organ doses and effective dose according to the ICRP Publication 110 reference male/female phantom and the modified ImPACT CT patient dosimetry.


ABSTRACT: We modified the Imaging Performance Assessment of CT scanners (ImPACT) to evaluate the organ doses and the effective dose based on the International Commission on Radiological Protection (ICRP) Publication 110 reference male/female phantom with the Aquilion ONE ViSION Edition scanner. To select the new CT scanner, the measurement results of the CTDI100,c and CTDI100,p for the 160 (head) and the 320 (body) mm polymethylmethacrylate phantoms, respectively, were entered on the Excel worksheet. To compute the organ doses and effective dose of the ICRP reference male/female phantom, the conversion factors obtained by comparison between the organ doses of different types of phantom were applied. The organ doses and the effective dose were almost identical for the ICRP reference male/female and m