An improvement in IMRT QA results and beam matching in linacs using statistical process control.
Ontology highlight
ABSTRACT: The purpose of this study is to apply the principles of statistical process control (SPC) in the context of patient specific intensity-modulated radiation therapy (IMRT) QA to set clinic-specific action limits and evaluate the impact of changes to the multileaf collimator (MLC) calibrations on IMRT QA results. Ten months of IMRT QA data with 247 patient QAs collected on three beam-matched linacs were retrospectively analyzed with a focus on the gamma pass rate (GPR) and the average ratio between the measured and planned doses. Initial control charts and action limits were calculated. Based on this data, changes were made to the leaf gap parameter for the MLCs to improve the consistency between linacs. This leaf gap parameter is tested monthly using a MLC sweep test. A follow-up dataset wit
SUBMITTER: Gagneur JD
PROVIDER: S-EPMC5711098 | biostudies-literature | 2014 Sep
REPOSITORIES: biostudies-literature