Dosimetry of gamma chamber blood irradiator using PAGAT gel dosimeter and Monte Carlo simulations.
Ontology highlight
ABSTRACT: Currently, the use of blood irradiation for inactivating pathogenic microbes in infected blood products and preventing graft-versus-host disease (GVHD) in immune suppressed patients is greater than ever before. In these systems, dose distribution and uniformity are two important concepts that should be checked. In this study, dosimetry of the gamma chamber blood irradiator model Gammacell 3000 Elan was performed by several dosimeter methods including thermoluminescence dosimeters (TLD), PAGAT gel dosimetry, and Monte Carlo simulations using MCNP4C code. The gel dosimeter was put inside a glass phantom and the TL dosimeters were placed on its surface, and the phantom was then irradiated for 5 min and 27 sec. The dose values at each point inside the vials were obtained from the magnetic reso