Measurement of gantry rotation time in modern ct.


ABSTRACT: The purpose of this study was to develop and evaluate a noninvasive method to assess rotation time in modern commercial computed tomography (CT) systems. The rotation time was measured at a selected nominal rotation time (400 ms) utilizing two types of solid-state detectors: the RTI's CT Dose Profiler (CTDP) and Unfors' Xi (Xi) probes. Either CTDP or Xi was positioned on the inner cover of the gantry and a sheet of lead (1 mm thick) placed on top of the detector. Since a pair of two successive peaks is used to determine the gantry rotation time, by necessity the helical scan must be employed. Upon completion of the data acquisition, these peak times were determined with the dedicated software to obtain rotation time. The average rotation time obtained with CTDP and Xi operated under the de