Comparison of planning techniques when air/fluid is present using the strut-adjusted volume implant (SAVI) for HDR-based accelerated partial breast irradiation.


ABSTRACT: The presence of air/fluid surrounding implantable devices used for partial breast irradiation may significantly impact dose coverage to at-risk tissue. Of the 67 total patients retrospectively evaluated for this study, 32 (48%) had greater than 1 cc volume of air/fluid extending outside of the strut-adjusted volume implant (SAVI) device surface and were selected for comparison of planning approaches. The planning approaches utilized two different definitions of PTV_EVAL. One definition of a PTV_EVAL (PTV_EVALSAVI) was based on expanding 1 cm beyond the SAVI device only while accounting for the air/fluid using the NSABP Protocol B-39/RTOG Protocol 0413. The second PTV_EVAL definition (PTV_EVALCAV) was based on expanding 1 cm beyond the cavity (SAVI device plus air/fluid volume). The results