Computed tomography dose index and dose length product for cone-beam CT: Monte Carlo simulations.


ABSTRACT: Dosimetry in kilovoltage cone beam computed tomography (CBCT) is a challenge due to the limitation of physical measurements. To address this, we used a Monte Carlo (MC) method to estimate the CT dose index (CTDI) and the dose length product (DLP) for a commercial CBCT system. As Dixon and Boone showed that CTDI concept can be applicable to both CBCT and conventional CT, we evaluated weighted CT dose index (CTDI(w)) and DLP for a commercial CBCT system. Two extended CT phantoms were created in our BEAMnrc/EGSnrc MC system. Before the simulations, the beam collimation of a Varian On-Board Imager (OBI) system was measured with radiochromic films (model: XR-QA). The MC model of the OBI X-ray tube, validated in a previous study, was used to acquire the phase space files of the full-fan and half

SUBMITTER: Kim S 

PROVIDER: S-EPMC5718669 | biostudies-literature | 2011 Jan

REPOSITORIES: biostudies-literature