Dosimetric characteristics of an electron multileaf collimator for modulated electron radiation therapy.
Ontology highlight
ABSTRACT: Modulated electron radiation therapy (MERT) has been proven as an effective way to deliver conformal dose distributions to shallow tumors while sparing distal critical structures and surrounding normal tissues. It had been shown that a dedicated electron multileaf collimator (eMLC) is necessary to reach the full potential of MERT. In this study, a manually-driven eMLC for MERT was investigated. Percentage depth dose (PDD) curves and profiles at different depths in a water tank were measured using ionization chamber and were also simulated using the Monte Carlo method. Comparisons have been performed between PDD curves and profiles collimated using the eMLC and conventional electron applicators with similar size of opening. Monte Carlo simulations were performed for all electron energies av
SUBMITTER: Eldib AA
PROVIDER: S-EPMC5719949 | biostudies-literature | 2010 Apr
REPOSITORIES: biostudies-literature
ACCESS DATA