Comparison of IPSM 1990 photon dosimetry code of practice with IAEA TRS-398 and AAPM TG-51.
Ontology highlight
ABSTRACT: Several codes of practice for photon dosimetry are currently used around the world, supported by different organizations. A comparison of IPSM 1990 with both IAEA TRS-398 and AAPM TG-51 has been performed. All three protocols are based on the calibration of ionization chambers in terms of standards of absorbed dose to water, as it is the case with other modern codes of practice. This comparison has been carried out for photon beams of nominal energies: 4 MV, 6 MV, 8 MV, 10 MV and 18 MV. An NE 2571 graphite ionization chamber was used in this study, cross-calibrated against an NE 2611A Secondary Standard, calibrated in the National Physical Laboratory (NPL). Absolute dose in reference conditions was obtained using each of these three protocols including: beam quality indices, beam quality c
SUBMITTER: Castrillon SV
PROVIDER: S-EPMC5720500 | biostudies-literature | 2009 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA