Design and implementation of a head-and-neck phantom for system audit and verification of intensity-modulated radiation therapy.
Ontology highlight
ABSTRACT: The head and neck is a challenging anatomic site for intensity-modulated radiation therapy (IMRT), requiring thorough testing of planning and treatment delivery systems. Ideally, the phantoms used should be anatomically realistic, have radiologic properties identical to those of the tissues concerned, and allow for the use of a variety of devices to verify dose and dose distribution in any target or normaltissue structure. A phantom that approaches the foregoing characteristics has been designed and built; its specific purpose is verification for IMRT treatments in the head-andneck region. This semi-anatomic phantom, HANK, is constructed of Perspex (Imperial Chemical Industries, London, U.K.) and provides for the insertion of heterogeneities simulating air cavities in a range of fixed posi
SUBMITTER: Webster GJ
PROVIDER: S-EPMC5721704 | biostudies-literature | 2008 Apr
REPOSITORIES: biostudies-literature
ACCESS DATA