Clinical laboratory verification of thyroglobulin concentrations in the presence of autoantibodies to thyroglobulin: comparison of EIA, radioimmunoassay and LC MS/MS measurements in an Urban Hospital.


ABSTRACT:

Objective

Thyroglobulin (Tg) measurements assess recurrence in post-thyroidectomy thyroid cancer patients. Tg measurements by enzyme immunoassays (EIA) can be falsely elevated by interference from Tg autoantibodies (TgAb). Radioimmunoassay (RIA) is less susceptible to TgAb interference and has been the standard-of-care test for TgAb positive patients. Recently developed liquid chromatography tandem mass spectrometry (LC-MS/MS) methods may eliminate TgAb interference. We assessed the performance of Tg measurements by EIA, RIA and LC-MS/MS to evaluate TgAb interference differences.

Results

We measured TgAb and Tg in 50 plasma samples from 40 patients in whom Tg measurement was part of their routine follow-up and 10 healthy volunteers. Discrepancy between EIA and both LC-MS/MS