Implementation of biological variation-based analytical performance specifications in the laboratory: Stringent evaluation of Improvacutor blood collection tubes.
Ontology highlight
ABSTRACT: Recently, because the quality of laboratory analyses has increased along with the need for quality improvement, several external quality control bodies have adapted performance specifications using the Desirable Biological Variation Database, termed "Ricos goals"; these criteria are more stringent than those presented in CLIA 88. In this study, we aimed to validate newly introduced serum separator tubes, Improvacutor, for routine clinical chemistry testing in accordance with Ricos goals and CLIA 88. Blood samples were collected from 100 volunteers into three types of serum vacuum tubes: Greiner Vacuette, Becton Dickinson (BD) Vacutainer, and Improve Improvacutor. The samples were subjected to 16 routine chemistry tests using a TBA-200fr NEO chemistry autoanalyzer. In the comparison analysi
SUBMITTER: Chung HJ
PROVIDER: S-EPMC5738124 | biostudies-literature | 2017
REPOSITORIES: biostudies-literature
ACCESS DATA