Unknown

Dataset Information

The impact of storage conditions upon gentamicin coated antimicrobial implants.


ABSTRACT: A systematic approach was developed to investigate the stability of gentamicin sulfate (GS) and GS/poly (lactic-co-glycolic acid) (PLGA) coatings on hydroxyapatite surfaces. The influence of environmental factors (light, humidity, oxidation and heat) upon degradation of the drug in the coatings was investigated using liquid chromatography with evaporative light scattering detection and mass spectrometry. GS coated rods were found to be stable across the range of environments assessed, with only an oxidizing atmosphere resulting in significant changes to the gentamicin composition. In contrast, rods coated with GS/PLGA were more sensitive to storage conditions with compositional changes being detected after storage at 60 °C, 75% relative humidity or exposure to light. The effect of γ-irradi

SUBMITTER: Mullins ND 

PROVIDER: S-EPMC5762933 | biostudies-literature | 2016 Dec

REPOSITORIES: biostudies-literature