Comparison of self-refraction using a simple device, USee, with manifest refraction in adults.


ABSTRACT:

Background

The USee device is a new self-refraction tool that allows users to determine their own refractive error. We evaluated the ease of use of USee in adults, and compared the refractive error correction achieved with USee to clinical manifest refraction.

Methods

Sixty adults with uncorrected visual acuity <20/30 and spherical equivalent between -6.00 and +6.00 diopters completed manifest refraction and self-refraction.

Results

Subjects had a mean (±SD) age of 53.1 (±18.6) years, and 27 (45.0%) were male. Mean (±SD) spherical equivalent measured by manifest refraction and self-refraction were -0.90 D (±2.53) and -1.22 diopters (±2.42), respectively (p = 0.001). The proportion of subjects correctable to ≥20/30 in the better eye was higher for manifest refraction (