Unknown

Dataset Information

Early aseptic loosening of a mobile-bearing total knee replacement.


ABSTRACT: Background and purpose - Registry-based studies have reported an increased risk of aseptic tibial loosening for the cemented Low Contact Stress (LCS) total knee replacement compared with other cemented designs; however, the reasons for this have not been established. We made a retrieval analysis with the aim of identifying the failure mechanism. Patients and methods - We collected implants, cement, tissue, blood, and radiographs from 32 failed LCS Complete cases. Damage to the tibial baseplate and insert was assessed. Exposure to wear products was quantified in 11 cases through analysis of periprosthetic tissue and blood. Implant alignment and bone cement thickness was compared with a control group of 43 non-revised cases. Results - Loosening of the tibial baseplate was the reason for revi

SUBMITTER: Kutzner I 

PROVIDER: S-EPMC5810837 | biostudies-literature | 2018 Feb