Optimized inversion-time schedules for quantitative T1 measurements based on high-resolution multi-inversion EPI.


ABSTRACT:

Purpose

Demonstrate an optimized multi-inversion echo-planar imaging technique to accelerate quantitative T1 mapping by judicious selection of inversion times for each slice.

Methods

Slice ordering is optimized to maximize discrimination between tissues with different T1 values. The optimized slice orderings are tested in the International Society for Magnetic Resonance in Medicine/National Institute of Standards and Technology phantom and compared with an unoptimized 21-measurement acquisition. The utility of the method is demonstrated in a healthy subject in vivo at 3 T and validated with a gold-standard inversion-recovery sequence. The in vivo precision of our technique was tested by repeated scans of the same subject within a scan session and across