ABSTRACT: Purpose
To characterize errors in enhancement in breast dynamic contrast-enhanced (DCE) MRI studies as a function of echo time and determine the source of dark band artifacts in clinical subtraction images.Methods
Computer simulations, oil and water substitute (methylene chloride), as well as an American College of Radiology quality control phantom were tested. Routine clinical DCE breast MRI study was bracketed with (accelerated) in-phase DCE acquisitions in five patients.Results
Simulation results demonstrated up to -160% suppression of the expected enhancement caused by differential enhancement of fat and water. Two-dimensional gradient-recalled echo and fat-suppressed 3D GRE phantom imaging confirmed the simulation results and showed that fat suppression does no