Use of esophageal balloon pressure-volume curve analysis to determine esophageal wall elastance and calibrate raw esophageal pressure: a bench experiment and clinical study.


ABSTRACT:

Background

Accurate measurement of esophageal pressure (Pes) depends on proper filling of the balloon. Esophageal wall elastance (Ees) may also influence the measurement. We examined the estimation of balloon-surrounding elastance in a bench model and investigated a simplified calibrating procedure of Pes in a balloon with relatively small volume.

Methods

The Cooper balloon catheter (geometric volume of 2.8 ml) was used in the present study. The balloon was progressively inflated in different gas-tight glass chambers with different inner volumes. Chamber elastance was measured by the fitting of chamber pressure and balloon volume. Balloon pressure-volume (P-V) curves were obtained, and the slope of the intermediate linear section was defined as the estimated chamber elastanc