Qualification of NISTmAb charge heterogeneity control assays.
Ontology highlight
ABSTRACT: The NISTmAb is a monoclonal antibody Reference Material from the National Institute of Standards and Technology; it is a class-representative IgG1κ intended serve as a pre-competitive platform for harmonization and technology development in the biopharmaceutical industry. The publication series of which this paper is a part describes NIST's overall control strategy to ensure NISTmAb quality and availability over its lifecycle. In this paper, the development and qualification of methods for monitoring NISTmAb charge heterogeneity are described. Capillary zone electrophoresis (CZE) and capillary isoelectric focusing (CIEF) assays were optimized and evaluated as candidate assays for NISTmAb quality control. CIEF was found to be suitable as a structural characterization assay yielding informat