Unknown

Dataset Information

0

Novel Compliant Scaffold with Specific Design for Venous System: Results of a Porcine Model Study.


ABSTRACT:

Background

Stenting has become the first-line treatment of obstructive venous disease because of poor results of balloon angioplasty. This preclinical study aimed to investigate the safety and efficacy profile of a novel compliant venous scaffold (CVS) denominated Petalo CVS, specifically designed for venous diseases.

Materials and methods

Twelve healthy pigs weighing 90 kg were used to test Petalo CVS. The devices were implanted into the internal jugular veins (IJVs) using a femoral vein percutaneous approach. The safety profile including the success rate of device releasing, anchoring, and positioning was evaluated immediately. Fracture, migration, primary patency, and endothelial response were assessed at 1, 2, 3, and 6 months after the study procedure.

Results

A total of 32 devices were successfully released in both IJVs. No procedure- or device-related complications were reported, and all pigs successfully completed the different scheduled follow-up periods. The primary patency rate was 100%, and no fracture or migration of the device into the brachiocephalic trunk was reported. Histological examination revealed only minimal lesions with minimal or absent inflammatory reaction surrounding the incorporated metallic rods.

Conclusions

This porcine model study showed a promising safety and efficacy profile of Petalo CVS, a novel endovenous device based on specific concepts.

SUBMITTER: Veroux P 

PROVIDER: S-EPMC5832035 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

Novel Compliant Scaffold with Specific Design for Venous System: Results of a Porcine Model Study.

Veroux Pierfrancesco P   Giaquinta Alessia A   Virgilio Carla C   Zani Davide Danilo DD   Ravasio Giuliano G   Ardita Vincenzo V   Secchiero Paola P   Scanziani Eugenio E   Zamboni Paolo P   Veroux Massimiliano M  

BioMed research international 20180131


<h4>Background</h4>Stenting has become the first-line treatment of obstructive venous disease because of poor results of balloon angioplasty. This preclinical study aimed to investigate the safety and efficacy profile of a novel compliant venous scaffold (CVS) denominated Petalo CVS, specifically designed for venous diseases.<h4>Materials and methods</h4>Twelve healthy pigs weighing 90 kg were used to test Petalo CVS. The devices were implanted into the internal jugular veins (IJVs) using a femo  ...[more]