Performance of a new real-time continuous glucose monitoring system: A multicenter pilot study.


ABSTRACT:

Aims/introduction

The present study aimed to investigate the performance of a new real-time continuous glucose monitoring system.

Materials and methods

Interstitial glucose levels were monitored for 7 days in 63 patients with type 1 or type 2 diabetes using the Medtrum A6 TouchCare® CGM System. Venous blood was collected on a randomized day of the wear period. Plasma glucose levels were measured as reference values.

Results

Among 1,678 paired sensor-reference values, 90.5% (95% confidence interval 89.1-91.9%) were within ±20%/20 mg/dL of the reference values, with a mean absolute relative difference of 9.1 ± 8.7% (95% CI: 8.9-9.2%). The percentages of paired sensor-reference values falling within zone A and B of the Clarke error grid analysis (EGA) and the