An Inexpensive, In-House-Made, Microdialysis Device for Measuring Drug-Protein Binding.
Ontology highlight
ABSTRACT: An inexpensive, in-house made microdialysis device is described that is suitable for measuring the binding of small molecules including drug candidates to serum proteins or other macromolecules. The device is based on the standard equilibrium dialysis method to measure the fraction of low molecular weight compound bound to proteins. It is constructed from a standard polypropylene 96-well plate, dialysis tubing, and low viscosity epoxy resin. The device can be readily prepared for a small fraction of the cost of a commercial, multichamber microdialysis device. Drug-protein binding results are provided, which validates the device.
SUBMITTER: Herbst ZM
PROVIDER: S-EPMC5846030 | biostudies-literature | 2018 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA