Is the dose distribution distorted in IMRT and RapidArc treatment when patient plans are swapped across beam-matched machines?
Ontology highlight
ABSTRACT: The purpose of this study is to evaluate the degree of dose distribution distortion in advanced treatments like IMRT and RapidArc when patient plans are swapped across dosimetrically equivalent so-called "beam-matched" machines. For this purpose the entire work is divided into two stages. At forefront stage all basic beam properties of 6 MV X-rays like PDD, profiles, output factors, TPR20/10 and MLC transmission of two beam-matched machines - Varian Clinac iX and Varian 600 C/D Unique - are compared and evaluated for differences. At second stage 40 IMRT and RapidArc patient plans from the pool of head and neck (H&N) and pelvis sites are selected for the study. The plans are swapped across the machines for dose recalculation and the DVHs of target and critical organs are evaluated for dose