Clinical experience with planning, quality assurance, and delivery of burst-mode modulated arc therapy.
Ontology highlight
ABSTRACT: "Burst-mode" modulated arc therapy (hereafter referred to as "mARC") is a form of volumetric-modulated arc therapy characterized by variable gantry rotation speed, static MLCs while the radiation beam is on, and MLC repositioning while the beam is off. We present our clinical experience with the planning techniques and plan quality assurance measurements of mARC delivery. Clinical mARC plans for five representative cases (prostate, low-dose-rate brain, brain with partial-arc vertex fields, pancreas, and liver SBRT) were generated using a Monte Carlo-based treatment planning system. A conventional-dose-rate flat 6 MV and a high-dose-rate non-flat 7 MV beam are available for planning and delivery. mARC plans for intact-prostate cases can typically be created using one 360° arc, and treatment