Commissioning and verification of the collapsed cone convolution superposition algorithm for SBRT delivery using flattening filter-free beams.


ABSTRACT: Linacs equipped with flattening filter-free (FFF) megavoltage photon beams are now commercially available. However, the commissioning of FFF beams poses challenges that are not shared with traditional flattened megavoltage X-ray beams. The planning system must model a beam that is peaked in the center and has an energy spectrum that is softer than the flattened beam. Removing the flattening filter also increases the maximum possible dose rates from 600 MU/min up to 2400 MU/min in some cases; this increase in dose rate affects the recombination correction factor, P(ion), used during absolute dose calibration with ionization chambers. We present the first reported experience of commissioning, verification, and clinical use of the collapsed cone convolution superposition (CCCS) dose calculati