CT and MR image fusion of tandem and ring applicator using rigid registration in intracavitary brachytherapy planning.
Ontology highlight
ABSTRACT: The purpose of this study is to find the uncertainties in the reconstruction of MR compatible ring-tandem intracavitary applicators of high-dose rate image-based brachytherapy treatment planning using rigid registration of 3D MR and CT image fusion. Tandem and ring reconstruction in MR image based brachytherapy planning was done using rigid registration of CT and MR applicator geometries. Verifications of registration for applicator fusion were performed in six verification steps at three different sites of tandem ring applicator set. The first site consists of three errors at the level of ring plane in (1) cranio caudal shift (Cranial Shift) of ring plane along tandem axis, (2) antero-posterior shift (AP Shift) perpendicular to tandem axis on the plane containing the tandem, and (3) later