Brachytherapy dose-volume histogram commissioning with multiple planning systems.


ABSTRACT: The first quality assurance process for validating dose-volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose-rate and high dose-rate radionuclide sources. A rectangular cuboid was contoured in five commercially available brachytherapy treatment planning systems. A single radioactive source commissioned for QA testing was positioned coplanar and concentric with one end. Using the brachytherapy dosimetry formalism defined in the AAPM Task Group 43 report series, calculations were performed to estimate dose deposition in partial volumes of the cuboid structure. The point-source approximation was used for a 125I source and the line-source approximation was used for a 192Ir source in simulated permanent and t

SUBMITTER: Gossman MS 

PROVIDER: S-EPMC5875493 | biostudies-literature | 2014 Mar