Sleep apnea detection by a cardiac resynchronization device integrated thoracic impedance sensor: A validation study against the gold standard polysomnography.


ABSTRACT:

Background

Sleep disordered breathing is a common but often undiagnosed comorbidity in heart failure patients. Cardiac implantable electronic devices used for cardiac resynchronization therapy (CRT) may detect sleep apnea by use of a transthoracic impedance sensor. Validation of the AP scan® algorithm (Boston Scientific®) was performed by using the diagnostic gold standard polysomnography (PSG).

Methods

Forty-one patients with impaired left ventricular ejection fraction, frequent right ventricular pacing due to atrioventricular block and heart failure symptoms despite optimal medical therapy underwent upgrading to biventricular pacing. Within one month after left ventricular lead implantation, sleep apnea was assessed by single-night PSG and AP scan® measurements.

Results