Double diffusion encoding MRI for the clinic.


ABSTRACT:

Purpose

The purpose of this study is to develop double diffusion encoding (DDE) MRI methods for clinical use. Microscopic diffusion anisotropy measurements from DDE promise greater specificity to changes in tissue microstructure compared with conventional diffusion tensor imaging, but implementation of DDE sequences on whole-body MRI scanners is challenging because of the limited gradient strengths and lengthy acquisition times.

Methods

A custom single-refocused DDE sequence was implemented on a 3T whole-body scanner. The DDE gradient orientation scheme and sequence parameters were optimized based on a Gaussian diffusion assumption. Using an optimized 5-min DDE acquisition, microscopic fractional anisotropy (μFA) maps were acquired for the first time in multiple sclerosis pa