Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry.
Ontology highlight
ABSTRACT: The aim of this study was to compare the axial lengths (ALs) using a new biometer with swept-source optical coherence tomography (Argos) versus ALs using a conventional biometer with partial coherence interferometry (IOL Master, version 5). The ALs in 48 eyes of 48 cataract patients were measured with Argos using refractive indexes that correspond to the particular tissue and with IOL Master using a single refractive index. The eyes were divided into three subgroups by AL length: short-AL group (n = 16), <23.27 mm; intermediate-AL group (n = 16), 23.27-24.03 mm; long-AL group (n = 16), ≥24.04 mm. The ALs (mm) measured with the Argos and IOL Master biometers, respectively, were 22.77 ± 0.43 and 22.74 ± 0.44, 23.63 ± 0.21 and 23.62 ± 0.21, and 26.00 ± 1.61 and 26.05 ± 1.64 in the short-, int
SUBMITTER: Higashiyama T
PROVIDER: S-EPMC5918161 | biostudies-literature | 2018
REPOSITORIES: biostudies-literature
ACCESS DATA