ABSTRACT: Background
Overinflation of the laryngeal mask airway (LMA) cuff may cause many of the complications associated with the use of the LMA. There is no clinically acceptable (cost effective and practical) method to ensure cuff pressure is maintained below the manufacturer's recommended maximum value of 60 cm H2O (44 mmHg). We studied the use of the intrinsic recoil of the LMA inflating syringe as an effective and practical way to limit cuff pressures at or below the manufacturer's recommended values.Methods
We enrolled 332 patients into three separate groups: LMAs inserted and inflated per standard practice at the institution with only manual palpation of the pilot balloon; LMA cuff pressures measured by a pressure transducer and reduced to < 60 cm H2O (4