An Intracardiac Soft Robotic Device for Augmentation of Blood Ejection from the Failing Right Ventricle.
Ontology highlight
ABSTRACT: We introduce an implantable intracardiac soft robotic right ventricular ejection device (RVED) for dynamic approximation of the right ventricular (RV) free wall and the interventricular septum (IVS) in synchrony with the cardiac cycle to augment blood ejection in right heart failure (RHF). The RVED is designed for safe and effective intracardiac operation and consists of an anchoring system deployed across the IVS, an RV free wall anchor, and a pneumatic artificial muscle linear actuator that spans the RV chamber between the two anchors. Using a ventricular simulator and a custom controller, we characterized ventricular volume ejection, linear approximation against different loads and the effect of varying device actuation periods on volume ejection. The RVED was then tested in vivo in adu
SUBMITTER: Horvath MA
PROVIDER: S-EPMC5937685 | biostudies-literature | 2017 Sep