Comparison of Flow Cytometric Methods for the Enumeration of Residual Leucocytes in Leucoreduced Blood Products: A Multicenter Study.
Ontology highlight
ABSTRACT: The BD FACSVia™ System features novel designs in hardware, software, and instrument QC. We compared the performance of the BD FACSVia System using the BD Leucocount™ kit with the BD FACSCalibur™ flow cytometer. Leucoreduced platelet (PLT, n = 252) and red blood cell (RBC, n = 278) specimens were enrolled at four sites. Each specimen was stained in four tubes using the BD Leucocount kit reagents and acquired on the two systems. BD Leucocount Control cells (high and low) were used to evaluate the inter-site reproducibility on the BD FACSVia System at three sites over 20 days. Deming regression and Bland-Altman analysis were performed to determine the WBC absolute counts on the BD FACSVia System vs. the BD FACSCalibur system. Assay accuracy for the range of 0-350 WBCs/µl was adequate. For sam
SUBMITTER: Zeng Y
PROVIDER: S-EPMC5947640 | biostudies-literature | 2018 Apr
REPOSITORIES: biostudies-literature
ACCESS DATA