ABSTRACT: Purpose
To report the findings and surgical techniques on two cases of severe tilting of an intraocular lens (IOL) that had been implanted with intrascleral fixation and the tilting was detected by anterior segment optical coherence tomography (AS-OCT).Observations
Two patients underwent flanged intrascleral fixation of an IOL with the double needle technique. AS-OCT showed that the tilt of IOL was 25.3° in Case 1 and was 38.1° in Case 2, and a second surgery was planned to reduce the IOL tilt. Both edges of the flanged haptics were externalized and shortened by 2-3 mm. Then, the haptics edges were inserted intrasclerally. The tilt of the IOLs was reduced to 7.7° and 5.7°, and the myopia-shifted refraction was reduced from -2.75 diopters (D) and -4.50 D to -0.13 D and -0.50