Unknown

Dataset Information

Calibration and performance of synchronous SIM/scan mode for simultaneous targeted and discovery (non-targeted) analysis of exhaled breath samples from firefighters.


ABSTRACT: Traditionally, gas chromatography-mass spectrometry (GC/MS) analysis has used a targeted approach called selected ion monitoring (SIM) to quantify specific compounds that may have adverse health effects. Due to method limitations and the constraints of preparing duplicate samples, the information that could be obtained from separately collecting the full scan chromatogram of the sample has often been sacrificed. However, the hybrid technique called synchronous SIM/scan mode alternates between the two acquisition modes, maintaining the accuracy and sensitivity of SIM for targeted analysis while also providing the full scan chromatogram for discovery of non-target compounds. This technology was assessed using calibration data and real-world breath samples from a joint EPA/NIOSH collaboration

SUBMITTER: Geer Wallace MA 

PROVIDER: S-EPMC5960982 | biostudies-literature | 2017 Sep

REPOSITORIES: biostudies-literature