Comparison of activated clotting times measured using the Hemochron Jr. Signature and Medtronic ACT Plus during cardiopulmonary bypass with acute normovolemic haemodilution.


ABSTRACT: Objective This study compared the activated clotting time (ACT) measured using the Hemochron Jr. Signature (HACT) with the ACT measured using the Medtronic ACT Plus (MACT) during cardiopulmonary bypass (CPB) with acute normovolemic haemodilution (ANH) in patients undergoing cardiac surgery. Methods The ACT was checked at baseline with both devices after inducing anaesthesia, and 400 to 800 mL of whole blood was withdrawn to induce moderate ANH. Before initiating CPB, a 300-IU/kg bolus dose of heparin was administered to maintain the HACT at >400 s; protamine was later given to reverse the anticoagulation. The ACT was checked using both devices at baseline, during heparinisation, and after protamine administration. Results In total, 106 pairs of samples from 29 patients were analysed. The A