Correlation between Insertion Torque and Implant Stability Quotient in Tapered Implants with Knife-Edge Thread Design.


ABSTRACT:

Aim

To evaluate the correlation between insertion torque (IT) and implant stability quotient (ISQ) in tapered implants with knife-edge threads.

Methods

Seventy-five identical implants (Anyridge, Megagen) were inserted by using a surgical drilling unit with torque control and an integrated resonance frequency analysis module (Implantmed, W&H). IT (N/cm) and ISQ were recorded and implants were divided into three groups (n = 25) according to the IT: low (<30), medium (30 < IT < 50), and high torque (>50). ISQ difference among groups was assessed by Kruskal-Wallis test, followed by Bonferroni-corrected Mann-Whitney U-test for pairwise comparisons. The strength of the association between IT and ISQ was assessed by Spearman Rho correlation coefficient (α = 0.05).

SUBMITTER: Baldi D 

PROVIDER: S-EPMC5976959 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature