Ontology highlight
ABSTRACT: Purpose
Magnetic resonance venography (MRV) has not been validated in pre-operative planning of the dural venous sinus stenting (VSS) among idiopathic intracranial hypertension (IIH) patients. We aim to prospectively evaluate dural venous sinus measurement in IIH patient population on two-dimensional time-of-flight (2D-TOF) MRV and Three-dimensional contrast-enhanced (3D-CE) MRV acquisitions and compare them against real-time endoluminal measurements with intravascular ultrasound (IVUS), served as the reference.Materials and methods
The study has been approved by the Weill Cornell Medicine institutional review board. All patients signed written informed consent approved by IRB. Prospective evaluation of forty-five consecutive IIH patients treated with VSS at our institution
SUBMITTER: Boddu SR
PROVIDER: S-EPMC5983447 | biostudies-literature | 2018
REPOSITORIES: biostudies-literature