Corneal epithelial permeability to fluorescein in humans by a multi-drop method.


ABSTRACT:

Purpose

The permeability of the corneal epithelium to fluorescein Pdc is an indicator of the health of the ocular surface. It can be measured in a clinical setting by determining the accumulation of fluorescein in the stroma following administration of the dye on the ocular surface. Here we demonstrate a new multi-drop method for the measurement of Pdc by a spot fluorometer.

Methods

Twenty-nine healthy participants were recruited for this study. First, a probe-drop of fluorescein (0.35%, 2 μL) was instilled on the conjunctiva. The clearance of the dye from the tears was immediately measured using the fluorometer. Following this, two loading drops (2%; 6 μL each) were administered 10 min apart. Fifteen minutes later, the ocular surface was washed and fluorescence from the str