A Robust Method for Ultrasound Beamforming in the Presence of Off-Axis Clutter and Sound Speed Variation.


ABSTRACT: Previously, we introduced a model-based beamforming algorithm to suppress ultrasound imaging artifacts caused by clutter sources, such as reverberation and off-axis scattering. We refer to this method as aperture domain model image reconstruction (ADMIRE). In this study, we evaluated the algorithm's limitations and ability to suppress off-axis energy using Field II-based simulations, experimental phantoms and in vivo data acquired by a Verasonics ultrasound system with a curvilinear transducer (C5-2). We compared image quality derived from a standard delay-and-sum (DAS) beamformer, DAS with coherence factor (CF) weighting, ADMIRE and ADMIRE plus CF weighting. Simulations, phantoms and in vivo scan results demonstrate that ADMIRE substantially suppresses off-axis energy, while preserving th