Dosimetric Implications of Computerised Tomography-Only versus Magnetic Resonance-Fusion Contouring in Stereotactic Body Radiotherapy for Prostate Cancer.


ABSTRACT: Background: Magnetic resonance (MR)-fusion contouring is the standard of care in prostate stereotactic body radiotherapy (SBRT) for target volume localisation. However, the planning computerised tomography (CT) scan continues to be used for dose calculation and treatment planning and verification. Discrepancies between the planning MR and CT scans may negate the benefits of MR-fusion contouring and it adds a significant resource burden. We aimed to determine whether CT-only contouring resulted in a dosimetric detriment compared with MR-fusion contouring in prostate SBRT planning. Methods: We retrospectively compared target volumes and SBRT plans for 20 patients treated clinically with MR-fusion contouring (standard of care) with those produced by re-contouring using CT data o