Ontology highlight
ABSTRACT: Goal
Develop and test an MRI-compatible hydrodynamic simulator of cerebrospinal fluid (CSF) motion in the cervical spinal subarachnoid space. Four anatomically realistic subject-specific models were created based on a 22-year-old healthy volunteer and a five-year-old patient diagnosed with Chiari I malformation.Methods
The in vitro models were based on manual segmentation of high-resolution T2-weighted MRI of the cervical spine. Anatomically realistic dorsal and ventral spinal cord nerve rootlets (NR) were added. Models were three dimensional (3-D) printed by stereolithography with 50-μm layer thickness. A computer controlled pump system was used to replicate the shape of the subject specific in vivo CSF flow measured by phase-contrast MRI. Each model was then scanned by T2
SUBMITTER: Thyagaraj S
PROVIDER: S-EPMC6025821 | biostudies-literature | 2018 Jul
REPOSITORIES: biostudies-literature