Beam focal spot position determination for an Elekta linac with the Agility® head; practical guide with a ready-to-go procedure.


ABSTRACT: A novel phantomless, EPID-based method of measuring the beam focal spot offset of a linear accelerator was proposed and validated for Varian machines. In this method, one set of jaws and the MLC were utilized to form a symmetric field and then a 180o collimator rotation was utilized to determine the radiation isocenter defined by the jaws and the MLC, respectively. The difference between these two isocentres is directly correlated with the beam focal spot offset of the linear accelerator. In the current work, the method has been considered for Elekta linacs. An Elekta linac with the Agility® head does not have two set of jaws, therefore, a modified method is presented making use of one set of diaphragms, the MLC and a full 360o collimator rotation. The modified method has been tested on tw

SUBMITTER: Chojnowski JM 

PROVIDER: S-EPMC6036348 | biostudies-literature | 2018 Jul