The reduction in treatment efficiency at high acoustic powers during MR-guided transcranial focused ultrasound thalamotomy for Essential Tremor.


ABSTRACT:

Purpose

To analyze clinical data indicating a reduction in the induced energy-temperature efficiency relationship during transcranial focused ultrasound (FUS) Essential Tremor (ET) thalamotomy treatments at higher acoustic powers, establish its relationship with the spatial distribution of the focal temperature elevation, and explore its cause.

Methods

A retrospective observational study of patients (n = 19) treated between July 2015 and August 2016 for (ET) by FUS thalamotomy was performed. These data were analyzed to compare the relationships between the applied power, the applied energy, the resultant peak temperature achieved in the brain, and the dispersion of the focal volume. Full ethics approval was received and all patients provided signed informed consent forms bef