ABSTRACT: Background
The software PRIMO for the Monte Carlo simulation of radiotherapy linacs could potentially act as a independent calculation system to verify the calculations of treatment planning systems. We investigated the suitability of the PRIMO default beam parameters to produce accurate dosimetric results for 6 MV photon beams from Varian Clinac 2100 linacs and 6 MV flattening-filter-free photon beams from Varian TrueBeam linacs.Methods
Simulation results with the DPM algorithm were benchmarked against a published reference dosimetry dataset based on point measurements of 25 dosimetric parameters on a large series of linacs. Studied parameters (for several field sizes and depths) were: PDD, off-axis ratios, and output factors for open fields and IMRT/SBRT-style fields. For
SUBMITTER: Hermida-Lopez M
PROVIDER: S-EPMC6081807 | biostudies-literature | 2018 Aug